Concept validation of a high dynamic range point-diffraction interferometer for wavefront sensing in adaptive optics

Nicolás Dubost*, Nazim Ali Bharmal, Marc Dubbeldam, Richard M. Myers

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

The direct detection and imaging of exoplanets requires the use of high-contrast adaptive optics (AO). In these systems quasi-static aberrations need to be highly corrected and calibrated. To achieve this, the pupil-modulated point-diffraction interferometer (m-PDI) was presented in an earlier paper. This present paper focuses on m-PDI concept validation through three experiments. First, the instrument’s accuracy and dynamic range are characterized by measuring the spatial transfer function at all spatial frequencies and at different amplitudes. Then, using visible monochromatic light, an AO control loop is closed on the system’s systematic bias to test for precision and completeness. In a central section of the pupil with 72% of the total radius, the residual error is 7.7 nm rms. Finally, the control loop is run using polychromatic light with a spectral FWHM of 77 nm around the R-band. The control loop shows no drop in performance with respect to the monochromatic case, reaching a final Strehl ratio larger than 0.7.
Original languageEnglish
Pages (from-to)4160-4167
Number of pages8
JournalApplied Optics
Volume61
Issue number14
Early online date25 Apr 2022
DOIs
Publication statusPublished - 9 May 2022

Fingerprint

Dive into the research topics of 'Concept validation of a high dynamic range point-diffraction interferometer for wavefront sensing in adaptive optics'. Together they form a unique fingerprint.

Cite this